Protein-Protein Variation of Protein Assays

Protein assays differ in their chemical basis for detecting protein-specific functional groups. Some assay methods detect peptide bonds, but no assay does this exclusively. Instead, each protein assay detects one or several different particular amino acids with greater sensitivity than than others. Consequently, proteins with different amino acid compositions produce color at different rates or intensities in any given protein assay.

The following table compares the protein-to-protein variability in color response of several Thermo Scientific Pierce Protein Assays. These data serve as a general guideline for evaluating response differences among protein samples. However, because the comparisons were made using one protein concentration and buffer, they should not be used as exact calibration factors.

This variability information is helpful for choosing a protein standard. For example, when the sample to be assayed is a purified antibody, bovine gamma globulin (BGG, protein #5) will be a more accurate standard than bovine serum albumin (BSA, protein #1). These data also indicate the importance of specifying which assay standard was used when reporting protein assay results.

Protein Methods Library Home

Learn more...

Overview of Protein Assays
Chemistry of Protein Assays

Related literature...

Protein Assay Handbook
Protein Assay Poster

View products...

BSA Protein Assay Standards
BGG Protein Assay Standards

Protein-to-protein variation of Thermo Scientific Pierce Protein Assays. For each of the protein assays presented here, 14 proteins were assayed using the standard test tube protocol. The net (blank corrected) average absorbance for each protein was calculated. The net absorbance for each protein is expressed as a ratio to the net absorbance for BSA (e.g., a ratio of 0.80 means that the protein produces 80% of the color obtained for an equivalent mass of BSA). All protein concentrations were at 1000µg/mL, except with the Micro BCA Assay which were at a concentration of 10µg/mL.
Results BCA
(Note 1)
Micro
BCA
Modified
Lowry
Coomassie
Plus
Coomassie
(Bradford)
Pierce
660nm
Relative Uniformity High High High Medium Low (Note 2) Low
Coefficient of Variation 14.7% 11.4% 11.9% 28.8% 38.2% 37%
Standard Deviation 0.15 0.12 0.13 0.21 0.26 0.27
Average ratio 1.02 1.05 1.09 0.73 0.68 0.74
Tested Protein ^ ^ ^ ^ ^ ^
1. Albumin, bovine serum 1.00 1.00 1.00 1.00 1.00 1.00
2. Aldolase, rabbit muscle 0.85 0.80 0.94 0.74 0.76 0.83
3. alpha-Chymotrypsinogen 1.14 0.99 1.17 0.52 0.48
4. Cytochrome C, horse heart 0.83 1.11 0.94 1.03 1.07 1.22
5. Gamma Globulin, bovine 1.11 0.95 1.14 0.58 0.56 0.51
6. IgG, bovine 1.21 1.12 1.29 0.63 0.58
7. IgG, human 1.09 1.03 1.13 0.66 0.63 0.57
8. IgG, mouse 1.18 1.23 1.20 0.62 0.59 0.48
9. IgG, rabbit 1.12 1.12 1.19 0.43 0.37 0.38
10. IgG, sheep 1.17 1.14 1.28 0.57 0.53
11. Insulin, bovine pancreas 1.08 1.22 1.12 0.67 0.60 0.81
12. Myoglobin, horse heart 0.74 0.92 0.90 1.15 1.19 1.18
13. Ovalbumin 0.93 1.08 1.02 0.68 0.32 0.54
14. Transferrin, human 0.89 0.98 0.92 0.90 0.84 0.8
15. a-Lactalbumin 0.82
16. Lysozyme 0.79
17. Trypsin inhibitor, soybean 0.38
Notes:
1. The BCA - Reducing Agent Compatible (BCA-RAC) Assay also produced a low coefficient of variation.
2. The Bio-Rad Bradford Protein Assay tested with the same proteins as our Coomassie (Bradford) Assay produced a very high coefficient of variation (46%), corresponding to very low relative uniformity.
Written and/or reviewed by Douglas Hayworth, Ph.D.

Instructions | MSDS | CofA
Product Instructions | MSDS | CofA  

Share This Page
 
 

Follow Us
  Email Sign-up  Email Announcements

 



Antibodies  |   Molecular Biology   |   Cell Biology   |  Thermo Scientific  |   * Trademarks  |   Privacy Statement
© 2014 Thermo Fisher Scientific Inc.   |   3747 N Meridian Rd, Rockford, IL USA 61101   |   1-800-874-3723  or  815-968-0747